Other, other text: this remediation mdr was generated under protocol (b)(4) , as a result of warning letter cms#(b)(4).The device was received for evaluation.Visual inspection showed labels were intact.Device was received in with damaged battery compartment.All blocks were found loose within the device.During functional check, the reported complaint was duplicated.The device was connected to an oxygen source and demand function testing was performed.Root cause was a faulty demand valve; valve was replaced.
|