SYNTHES GMBH 2.5MM DRILL BIT/QC/GOLD/110MM; INSTRUMENT, SURGICAL, ORTHOPEDIC, PNEUMATIC POWERED & ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
Catalog Number 310.250 |
Device Problems
Break (1069); Entrapment of Device (1212)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 12/13/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Additional narrative: additional device product codes: gfa; gff; hwe.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that the drill bit broke during the procedure on (b)(6) 2021 and affected the patient, as the tip remained inside the patient.This report is for one (1) 2.5mm drill bit/qc/gold/110mm.This is report 1 of 1 for complaint (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: part 310.250, lot 69p1354: manufacturing site: bettlach.Release to warehouse date: september 16, 2020.A manufacturing record evaluation was performed for the finished device lot and no non-conformance was identified.H3, h6: a photo investigation was completed: the device was not returned.A photo-investigation was performed on the images provided.Upon inspecting the images provided, the drill bit was observed to be broken, reported condition can be confirmed, the broken portion of the device is not shown on the picture.There is no evidence provided to confirm the embedded condition reported.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.During the investigation, no product design issues, or discrepancies were observed.No manufacturing issues were noted during investigation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Therefore, it has been determined that no corrective and/or preventive action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was further reported the procedure was successfully completed with no delay and medical intervention needed.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Additionally, a visual inspection was conducted of the returned device from the received images.Visual analysis of the returned sample revealed that the drill bit ø2.5 l110/85 2flute f/qc was found to be broken from the tip; fragment is not included in the package.Embedded device condition cannot be confirmed since x-ray evidence was not provided.A dimensional inspection for the drill bit ø2.5 l110/85 2flute f/qc was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces as mentioned in the event description.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the drill bit ø2.5 l110/85 2flute f/qc would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed the following drawings reflecting the current and manufactured revisions were reviewed: drill bit 2-flute dx.X - d4.4 drill bit for quick coupling.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|