• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS SCARBOROUGH, INC BINAXNOW COVID-19 ANTIGEN SELF TEST; LATERAL FLOW IMMUNOASSAY IVD OF COVID-19

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS SCARBOROUGH, INC BINAXNOW COVID-19 ANTIGEN SELF TEST; LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 Back to Search Results
Catalog Number 195-160
Device Problem Use of Device Problem (1670)
Patient Problems Red Eye(s) (2038); Eye Burn (2523)
Event Date 12/14/2021
Event Type  Injury  
Manufacturer Narrative
This investigation is still in progress.Once the investigation is completed, a supplemental report will be provided.
 
Event Description
The consumer reported that the extraction reagent from the binaxnow covid-19 antigen self test came in contact with the eyes on (b)(6) 2021.The consumer spoke with technical services (ts) and stated that he rinsed his eyes for five minutes, the consumer stated he was experiencing eye redness and stinging.Ts connected the patient with a local poison and drug services line for further medical advice on how to proceed.Due to privacy reasons ts had to release the call once the consumer was connected to a specialist at the local poison and drug services center.No additional patient information, including treatment and outcome, was provided.
 
Manufacturer Narrative
Technical services followed up with the customer and they confirmed they received the sds.They stated their eyes are fine and they are not carrying the reagent in their pocket to avoid this mistake from occurring again.According to the package insert in195150c v.3.0: precautions 20.The reagent solution contains a harmful chemical.If the solution contacts the skin or eye, flush with copious amounts of water.If irritation persists, seek medical advice.Based on the above summary, the investigation is deemed complete.The product will continue to be monitored and tracked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BINAXNOW COVID-19 ANTIGEN SELF TEST
Type of Device
LATERAL FLOW IMMUNOASSAY IVD OF COVID-19
Manufacturer (Section D)
ABBOTT DIAGNOSTICS SCARBOROUGH, INC
10 southgate road
scarborough ME 04074
Manufacturer Contact
rachel blackwell
10 southgate road
scarborough, ME 04074
6613888803
MDR Report Key13211802
MDR Text Key283576582
Report Number1221359-2022-00018
Device Sequence Number1
Product Code QKP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EUA210264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Catalogue Number195-160
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/15/2021
Initial Date FDA Received01/10/2022
Supplement Dates Manufacturer Received04/15/2022
Supplement Dates FDA Received05/10/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-