Model Number CLV-190
Device Problem Excessive Heating (4030)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
A field service engineer (fse) was onsite to troubleshoot the issue and the reported e101 error was confirmed.The fse performed general operation and safety checks with positive results.In addition, the fse found dust that blocked the air flow of the fan and the dust was eliminated with compressed air.Lastly, the xenon lamp maj-1817 was replaced.The device will not be returned for further evaluation and repair.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer initiated an onsite repair request because the evis exera iii xenon light source displayed an e101 error (overheating).The event occurred after procedure and there was no patient harm or user injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, it is likely that this phenomenon occurred because the fan supply was cut off due to a large amount of dust inside the airframe.The instruction manual identifies the following related verbiage that may help prevent the phenomenon: ¿avoid using the light source in a dusty environment.This may damage the light source.Clean and vacuum dust the ventilation grills using a vacuum cleaner.Otherwise, the light source may break down and gets damaged from overheating.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13213120
MDR Text Key291965669
Report Number8010047-2022-01155
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/13/2021
Initial Date FDA Received01/10/2022
Supplement Dates Manufacturer Received01/25/2022
Supplement Dates FDA Received02/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/14/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
XENON LAMP MAJ-1817
-
-