Model Number ACL TIGHTROPE RT |
Device Problems
Break (1069); Disconnection (1171); Loose or Intermittent Connection (1371)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/13/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
|
Event Description
|
It was reported that during an anterior cruciate ligament surgery the tightrope thread broke off while tensioning the graft in the femur after the screw placement in the tibia.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.25-dec-2021 update: further information to clarify the event was provided.It was reported that the implant (graft) got disconnected from the tightrope treads after the tightrope treads got cut and it got loose.Therefore the femoral fixation was lost.The surgeon had to remove the interference screw from the tibia and then remove the entire graft and replace the cut tightrope rt with a new one.Finally the surgeon replaced also the damaged bio-composite interference screw with another peek interference screw.
|
|