ARTHREX, INC. ACL TIGHTROPE RT; PIN, FIXATION, SMOOTH Back to Search Results
Model Number ACL TIGHTROPE RT
Device Problems Break (1069); Disconnection (1171); Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/13/2021
Event Type  Injury  
Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event Description
It was reported that during an anterior cruciate ligament surgery the tightrope thread broke off while tensioning the graft in the femur after the screw placement in the tibia.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.25-dec-2021 update: further information to clarify the event was provided.It was reported that the implant (graft) got disconnected from the tightrope treads after the tightrope treads got cut and it got loose.Therefore the femoral fixation was lost.The surgeon had to remove the interference screw from the tibia and then remove the entire graft and replace the cut tightrope rt with a new one.Finally the surgeon replaced also the damaged bio-composite interference screw with another peek interference screw.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACL TIGHTROPE RT
Type of Device
PIN, FIXATION, SMOOTH
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key13225885
MDR Text Key283668085
Report Number1220246-2022-04284
Device Sequence Number1
Product Code HTY
UDI-Device Identifier00888867019553
UDI-Public00888867019553
Combination Product (y/n)N
Reporter Country CodeIZ
PMA/PMN Number
K112990
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 01/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberACL TIGHTROPE RT
Device Catalogue NumberAR-1588RT
Device Lot Number10370691
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/25/2021
Initial Date FDA Received01/11/2022
Date Device Manufactured09/06/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-