BOSTON SCIENTIFIC CORPORATION VALITUDE X4 CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number U128 |
Device Problems
False Positive Result (1227); Pacing Problem (1439); Under-Sensing (1661)
|
Patient Problem
Atrial Fibrillation (1729)
|
Event Date 09/16/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
|
Event Description
|
It was reported that review of the stored data from this device identified signal artifact monitor events which were false positives due to atrial fibrillation.Additionally, atrial undersensing resulted in pacing into the arrhythmia which was conducted into the right ventricle for more than two minutes at a rate of 235 bpm.A boston scientific technical services consultant discussed options to resolve the clinical observations.The device was subsequently explanted.No additional adverse patient effects were reported.
|
|
Event Description
|
It was reported that review of the stored data from this device identified signal artifact monitor events which were false positives due to atrial fibrillation.Additionally, atrial undersensing resulted in pacing into the arrhythmia which was conducted into the right ventricle for more than two minutes at a rate of 235 bpm.A boston scientific technical services consultant discussed options to resolve the clinical observations.The device was subsequently explanted.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
|
Search Alerts/Recalls
|
|
|