Model Number CYF-5 |
Device Problems
Loose or Intermittent Connection (1371); Defective Component (2292)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/14/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device inspection found the forceps channel port is loose, due to looseness of forceps channel port, water tightness is lost and air leak inspection fail.The b-port needs to be replaced.Based on device evaluation findings the reported failure was confirmed.Investigation is ongoing.This report will be supplemented accordingly following investigation.
|
|
Event Description
|
As reported, the device flushing port is loose.The issue found during reprocessing.There is no patient involvement associated on this reported event.No user injury reported.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been less than 1 year since the subject device was manufactured.Based on the results of the investigation, it is likely the biopsy channel port of the subject device received excessive force while being used with maj-891 forceps/irrigation plug.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
|
|