• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AUTOGEN; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AUTOGEN; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number D177
Device Problems High impedance (1291); Failure to Sense (1559); High Capture Threshold (3266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2021
Event Type  malfunction  
Manufacturer Narrative
Device remains in service.As no further information concerning this report is expected our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
It was reported that this implantable cardioverter defibrillator (icd) device exhibited increasingly high pacing impedance from 999 ohms to 1,780 ohms, rising thresholds 3.1v@1.2msec, and failure to sense.The device remains implanted.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOGEN
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13233042
MDR Text Key283664883
Report Number2124215-2021-39289
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/08/2016
Device Model NumberD177
Device Catalogue NumberD177
Device Lot Number371024
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2021
Initial Date FDA Received01/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/27/2014
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age81 YR
Patient SexMale
-
-