Model Number GIF-Q260J |
Device Problem
Mechanical Problem (1384)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The subject device was returned to olympus service operation repair center (sorc).Sorc checked the subject device and found that the reported phenomenon was duplicated.There was the possibility that insufficiently or incorrectly reprocessed subject device might be used for next procedure.The exact cause has been under investigation.Therefore, the exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
|
|
Event Description
|
Olympus medical systems corp.(omsc) was informed from the user that during unspecified timing, it was found that the suction valve could not be detached from the subject device.The occurrence date of the event is unknown, and there was no report of patient injury associated with this event.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to withdraw mfr report # 8010047-2022-01303.Olympus medical systems corp.(omsc) confirmed that there was no mdr reportable malfunction or adverse event.
|
|