MAUDE Adverse Event Report: ZIMMER BIOMET, INC. M2A-MAGNUM PF CUP 52ODX46ID; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. M2A-MAGNUM PF CUP 52ODX46ID; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Material Erosion (1214)
Patient Problems Pain (1994); Burning Sensation (2146); Metal Related Pathology (4530); Solid Tumour (4552); Swelling/ Edema (4577)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Implant date: unknown month and day in 2006.The device will not be returned for analysis, as the device remains implanted; however, an investigation of the reported event is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2022-00087, 0001825034-2022-00088.
 
Event Description
It was reported that approximately 15 years post implantation, the patient is experiencing severe burning, pain, and massive swelling due to the metal on metal hip that is causing swelling of her joints.Orthopedic surgeon has aspirated to remove the metal debris/blood collection.The surgeon recently told her after having recent surgery to partially remove a massive tumor that is being caused by the hip that her anatomy is being eaten away from the metal.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated:.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Radiograph reports were provided and reviewed by a health care professional.Review of the available records identified the following: x-ray, mri, ct, fluoroscopy report findings of status post right hip arthroplasty.Intact acetabular and femoral components with effusion, pseudocapsule, synovitis and metal debris.Ct and doppler of right thigh findings are 2 fluid collections in right pelvis and right thigh.The larger fluid structure in the right pelvis extends inferiorly into the soft tissues of the right thigh, probable myomatous of the uterus.The right hip structure is suggestive of a cyst.Doppler findings, no dvt, patent venous structures.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4, b5, d4, g3, g6, h2, h3, h4, h10 corrected: d4 (lot #).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M2A-MAGNUM PF CUP 52ODX46ID
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13237622
MDR Text Key283686720
Report Number0001825034-2022-00086
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101336
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 03/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2016
Device Model NumberN/A
Device Catalogue NumberUS157852
Device Lot Number023280
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/21/2021
Initial Date FDA Received01/12/2022
Supplement Dates Manufacturer Received02/03/2022
02/24/2022
03/28/2022
Supplement Dates FDA Received02/24/2022
03/21/2022
03/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
UNKNOWN HEAD; UNKNOWN TAPER ADAPTER
Patient Outcome(s) Other;
Patient SexFemale
-
-