MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE CONNECTED OR CART, 120 V; CART, EMERGENCY, CARDIOPULMONARY (EXCLUDING EQUIPMENT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE CONNECTED OR CART, 120 V; CART, EMERGENCY, CARDIOPULMONARY (EXCLUDING EQUIPMENT) Back to Search Results
Model Number 240099155
Device Problem Arcing (2583)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/10/2021
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that there was arcing between the cables.
 
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that there was arcing between the cables.
 
Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.The reported failure mode will be monitored for future reoccurrence.Alleged failure: arcing cables probable root cause: manufacturing nonconformance, castor design, overloaded castor, castor damage from cleaning products, base damage, damage during shipping replacement parts, use error.Manufacture date is not known.
 
Event Description
It was reported that there was arcing between the cables.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONNECTED OR CART, 120 V
Type of Device
CART, EMERGENCY, CARDIOPULMONARY (EXCLUDING EQUIPMENT)
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key13237824
MDR Text Key285986930
Report Number0002936485-2022-00040
Device Sequence Number1
Product Code BZN
UDI-Device Identifier00858701006360
UDI-Public00858701006360
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number240099155
Device Catalogue Number0240099155
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/15/2021
Initial Date FDA Received01/12/2022
Supplement Dates Manufacturer Received12/15/2021
Supplement Dates FDA Received03/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-