• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. RSP; GLENOID, HEAD W/RETAINING SCREW, RSP, 36MM, NEUTRAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. RSP; GLENOID, HEAD W/RETAINING SCREW, RSP, 36MM, NEUTRAL Back to Search Results
Model Number 508-36-101
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem Pain (1994)
Event Date 12/21/2021
Event Type  Injury  
Manufacturer Narrative
The reason for this revision surgery was reported as pain.The previous surgery and the surgery detailed in this event occurred 5 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There was an ncmr# (b)(4) associated with the main part #508-36-101,glenoid, head w/retaining screw, rsp, 36mm, neutral which document that out of 20 parts lot 1 item was rejected due to operator error outside of first piece setup.Later, the rejected parts were reworked and accepted after proper justification.All other items in the lot were met with fit, form and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to pain.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event Description
Revision surgery: patient is having posterior pain when he goes through the range of motion.Surgeon said he wants to change the lateralization of the glenosphere (different size) also a different thickness of polyethylene on the humeral implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP
Type of Device
GLENOID, HEAD W/RETAINING SCREW, RSP, 36MM, NEUTRAL
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
kiersten soderman
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key13237924
MDR Text Key284167755
Report Number1644408-2022-00035
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888912024761
UDI-Public(01)00888912024761
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number508-36-101
Device Catalogue Number508-36-101
Device Lot Number869C2960
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/21/2021
Initial Date FDA Received01/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/06/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
509-00-036 LOT 378P1379; 510-08-000 LOT 142G1393
Patient Outcome(s) Required Intervention;
Patient Age53 YR
Patient SexMale
-
-