BOSTON SCIENTIFIC CORPORATION EASYTRAK 2 IS-1; IMPLANTABLE LEAD Back to Search Results
Model Number 4542
Device Problems Failure to Capture (1081); Fracture (1260); Impedance Problem (2950); High Capture Threshold (3266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2021
Event Type  Injury  
Event Description
It was reported that this left ventricular (lv) lead appears to be fractured, as it exhibited loss of capture (loc) and high impedances noticed during routine check.The plan is to just monitor as the device is currently programmed to only rv and not pacing.No additional adverse patient effects were reported.
 
Event Description
It was reported that this left ventricular (lv) lead appears to be fractured, as it exhibited loss of capture (loc) and high impedances noticed during routine check.The plan is to monitor as the device is currently programmed to only right ventricular (rv) and not pacing.No additional adverse patient effects were reported.Product remains in service.Additional information was received which indicated that this lv lead was explanted and replaced due to high pacing thresholds cause by a fracture.The lv lead will not be returned.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2 IS-1
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13262257
MDR Text Key285476080
Report Number2124215-2021-39895
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526410901
UDI-Public00802526410901
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P010012/S024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/04/2015
Device Model Number4542
Device Catalogue Number4542
Device Lot Number161291
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2021
Initial Date FDA Received01/14/2022
Supplement Dates Manufacturer Received07/10/2023
Supplement Dates FDA Received07/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/04/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age66 YR
Patient SexFemale
-
-