Model Number 4542 |
Device Problems
Failure to Capture (1081); Fracture (1260); Impedance Problem (2950); High Capture Threshold (3266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/07/2021 |
Event Type
Injury
|
Event Description
|
It was reported that this left ventricular (lv) lead appears to be fractured, as it exhibited loss of capture (loc) and high impedances noticed during routine check.The plan is to just monitor as the device is currently programmed to only rv and not pacing.No additional adverse patient effects were reported.
|
|
Event Description
|
It was reported that this left ventricular (lv) lead appears to be fractured, as it exhibited loss of capture (loc) and high impedances noticed during routine check.The plan is to monitor as the device is currently programmed to only right ventricular (rv) and not pacing.No additional adverse patient effects were reported.Product remains in service.Additional information was received which indicated that this lv lead was explanted and replaced due to high pacing thresholds cause by a fracture.The lv lead will not be returned.No additional adverse patient effects were reported.
|
|