BOSTON SCIENTIFIC CORPORATION EASYTRAK 2 IS-1; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number 4543 |
Device Problems
Signal Artifact/Noise (1036); Failure to Capture (1081); Fracture (1260); High impedance (1291); Failure to Read Input Signal (1581); High Capture Threshold (3266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/20/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
The local area field representative was contacted for additional information.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that this left ventricular (lv) lead triggered alerts due to out of range intrinsic amplitude and high out-of-range pace impedance measurements greater than 3,000 ohms.In-clinic testing revealed variable lv pace impedances from 1,100 to >2,500 ohms.Lv thresholds also varied from.8v@0.5ms to non-capture at 6v@0.5ms.The lv lead was programmed to true bipolar, and it was noted that the patient was not pacer dependent.Noise was observed during isometric testing.The field representative was instructed to reprogram the lead from tip to can and check impedances and thresholds; pace impedances were consistently >2,500 ohms and threshold was 2.5v@0.5ms.Technical services (ts) discussed that this was likely a lead fracture with intermittent make-break connection that allowed for occasional in-range measurements.This lv lead remains in service.No adverse patient effects were reported.
|
|
Event Description
|
It was reported that this left ventricular (lv) lead triggered alerts due to out of range intrinsic amplitude and high out-of-range pace impedance measurements greater than 3,000 ohms.In-clinic testing revealed variable lv pace impedances from 1,100 to >2,500 ohms.Lv thresholds also varied from.8v@0.5ms to non-capture at 6v@0.5ms.The lv lead was programmed to true bipolar, and it was noted that the patient was not pacer dependent.Noise was observed during isometric testing.The field representative was instructed to reprogram the lead from tip to can and check impedances and thresholds; pace impedances were consistently >2,500 ohms and threshold was 2.5v@0.5ms.Technical services (ts) discussed that this was likely a lead fracture with intermittent make-break connection that allowed for occasional in-range measurements.This lv lead remains in service.No adverse patient effects were reported.Additional information received reported that the patient was seen in clinic for further evaluation.A better vector was found lvring to rv, and consistent measurements were obtained in office and again via remote transmission.Rv pace impedance measurement was 685 ohms via a recent remote transmission.It was noted that autocapture was not on.This lv lead remains in service.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that this left ventricular (lv) lead triggered alerts due to out of range intrinsic amplitude and high out-of-range pace impedance measurements greater than 3,000 ohms.In-clinic testing revealed variable lv pace impedances from 1,100 to >2,500 ohms.Lv thresholds also varied from.8v@0.5ms to non-capture at 6v@0.5ms.The lv lead was programmed to true bipolar, and it was noted that the patient was not pacer dependent.Noise was observed during isometric testing.The field representative was instructed to reprogram the lead from tip to can and check impedances and thresholds; pace impedances were consistently >2,500 ohms and threshold was 2.5v@0.5ms.Technical services (ts) discussed that this was likely a lead fracture with intermittent make-break connection that allowed for occasional in-range measurements.This lv lead remains in service.No adverse patient effects were reported.Additional information received reported that the patient was seen in clinic for further evaluation.A better vector was found lvring to rv, and consistent measurements were obtained in office and again via remote transmission.Rv pace impedance measurement was 685 ohms via a recent remote transmission.It was noted that autocapture was not on.This lv lead remains in service.No adverse patient effects were reported.Additional information received reported that this left ventricular (lv) lead was explanted and replaced.It was noted that the cardiac resynchronization therapy defibrillator (crt-d) was also explanted and replaced during this procedure due to normal battery depletion (nbd).No additional adverse patient effects were reported.
|
|
Search Alerts/Recalls
|
|
|