It was reported that the right ventricular (rv) lead was initially thought to have insulation damage due to low pacing impedances, however additional testing was performed including an x-ray that determined that this lead was dislodged into the right atrium/high right ventricle due to twiddler syndrome of this device.The patient was hospitalized due to observations of loss of capture, inappropriate/ineffective pacing and sensing issues in order to be monitored externally until a revision can be performed.Subsequently, the device was explanted and replaced, however the rv lead was isolated and unable to be explanted so surgically capped and replaced.No additional adverse patient effects were reported.
|