Catalog Number 6393350
Device Problems Material Opacification (1426); Material Protrusion/Extrusion (2979)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/20/2020
Event Type  malfunction  
Event Description
It was reported that approximately one year post catheter placement, the device allegedly had white opacification of the transparent part.There was no reported patient injury.
 
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, photos were provided for review.The investigation of the reported event is currently underway.(expiration date: 05/2021).
 
Event Description
It was reported that sometime post catheter placement, the device allegedly had white opacification of the transparent part.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one 35cm glidepath d/l catheter was returned for evaluation.Gross visual, microscopic visual and functional evaluations were performed on the returned device.The investigation is confirmed for the reported material opacification and identified material protrusion issues as the extension legs of the catheter was noted to be bulged and appeared emulsified in a milky white color just proximal to the bifurcation.Further during functional evaluation, the catheter was patent to infusion and aspiration without any issue.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 05/2021), g3, h6 (device) h11: b5, h6 (method, result, conclusion) h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDEPATH
Type of Device
DIALYSIS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13287697
MDR Text Key284001563
Report Number3006260740-2021-05666
Device Sequence Number1
Product Code MSD
UDI-Device Identifier00801741138539
UDI-Public(01)00801741138539
Combination Product (y/n)N
Reporter Country CodeSA
PMA/PMN Number
K051748
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6393350
Device Lot NumberREDV1566
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/23/2021
Initial Date FDA Received01/19/2022
Supplement Dates Manufacturer Received04/18/2022
Supplement Dates FDA Received04/19/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
Patient Weight70 KG
-
-