• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AESPIRE 7100; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AESPIRE 7100; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2021
Event Type  malfunction  
Manufacturer Narrative
A leak condition will be noted in the preop check of the equipment, as contained in the user manual.The leak may be able to be compensated for or made up with fresh gas flow.However this failure was discovered during patient use, which may mean the canister had to be replaced during the procedure.The biomedical technician noticed the cracked canister lid during a review of the unit after the case.The manufacturer was contacted for replacement.Patient information could not be obtained due to country privacy laws.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Legal manufacturer: (b)(4).Patient information could not be obtained due to country privacy laws.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Legal manufacturer: (b)(4).
 
Event Description
Per (b)(6) aer database: it was reported the absorbent canister would not latch to the unit during patient use.The anesthesia unit was replaced and the surgery was completed without injury or harm to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESPIRE 7100
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
anthony amenson
3030 ohmeda drive
madison, WI 53718
MDR Report Key13307240
MDR Text Key287259473
Report Number2112667-2022-00137
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/23/2021
Initial Date FDA Received01/20/2022
Date Device Manufactured07/28/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-