The sensor was investigated.Review of in-vivo data in dms shows that ec#30 (led disconnect error) was triggered on (b)(6) 2021.In-vitro qc test did not find any performance issue with the sensor.However, sensor manager software measurements test of the sensor showed that the led flag for sensor (b)(6) cleared at field current threshold higher than 412 adc counts which is the root cause of this issue.As part of resolution, the rma was authorized to offer the user a sensor replacement.D4: udi number updated to (b)(4).D9: device available for evaluation? yes, 02 february 2022.H3: device evaluated by manufacturer? yes.H6: type of investigation updated to 10.H6: investigation findings updated to 4203.H6: investigation conclusions updated to 4307.
|