• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number B1070-040
Device Problems Difficult to Remove (1528); Material Rupture (1546); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/27/2021
Event Type  Injury  
Manufacturer Narrative
The customer reported the device is not returning.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.The other armada balloon is filed under a separate medwatch report number.
 
Event Description
It was reported that the procedure was to treat a left femoral arteriovenous (av) graft.An 8x80mm armada 35 and 7x40mm armada 35 balloon dilatation catheter (bdc) were being used in the procedure and advanced separately to the av graft.Both balloons were inflated two to three times each, to 13 atmospheres (atm) when the balloons ruptured.In an attempt to pull the balloons back into the sheath for removal, a separation occurred.Some of the balloon pieces were removed from the anatomy with endoscopic forceps.The patient¿s entire anatomy was x-rayed to try to find the remaining broken portion of the balloon, but it was unable to be found.The patient came back to the operating room on the (b)(6) 2021; however, the outcome from that procedure is unknown.It was noted that ¿it is less likely that the balloons ruptured, but instead were snagged on stents that were already in place in the patient¿s leg¿.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined that the reported difficulties and subsequent treatments were likely due to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Type of Device
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13311602
MDR Text Key284516762
Report Number2024168-2022-00718
Device Sequence Number1
Product Code LIT
UDI-Device Identifier08717648154782
UDI-Public08717648154782
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberB1070-040
Device Catalogue NumberB1070-040
Device Lot Number10920G2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/28/2021
Initial Date FDA Received01/20/2022
Supplement Dates Manufacturer Received02/17/2022
Supplement Dates FDA Received02/23/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age48 YR
Patient SexMale
Patient Weight116 KG
-
-