Model Number CLV-S45
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
The subject device was inspected at olympus service operation repair center (sorc).Sorc checked the subject device and found that the reported phenomenon was duplicated and the error of turret motor fatigue occurred.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the user that during the unspecified timing, the front panel display flashed and could not be operated.The occurrence date of the event is unknown.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.However, based on the reported information, it is presumed that the phenomenon is due to fatigue of the turret motor.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13333296
MDR Text Key295419870
Report Number8010047-2022-01851
Device Sequence Number1
Product Code NWB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S45
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/27/2021
Initial Date FDA Received01/24/2022
Supplement Dates Manufacturer Received02/03/2022
Supplement Dates FDA Received03/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/09/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-