Catalog Number MX4810P1CZ
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/1901
Event Type  malfunction  
Event Description
It was reported that air escapes from the bag and is therefore leaking, no injury occurred.
 
Manufacturer Narrative
One unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.Dhr review was done, no issues related to the original complaint were found.
 
Manufacturer Narrative
Other, other text: this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No problem or issues were identified during the device history record review.Visual inspection showed some signs of damage - scratches on the cuff which may correspond to use of the product, crumpled tubing from forceps usage, broken pressure gauge without coverslip for sample (probably caused by improper use).During functional testing, the sample was inflated to 300mm hg and the c-fusor kept the pressure for 2 minutes.Both samples stayed inflated, no leaking from the bag was observed.Based on the first investigational results, the reported issues could not be confirmed.As per the customer a second evaluation was performed.Products were inflated to 300mm hg and hung up for observation.After about thirty minutes, the pressure began to drop slightly for sample.After two hours, a pressure drop of about 50 mm hg was observed.No pressure drop was observed for the second sample.Another testing of the first sample was performed.To identify the place of leakage, the inflated sample was put in the water.Leakage was observed between luer lock male and stopcock 4-way white.These two parts are joined manually by tightening.However, even after proper tightening, the leakage was still observed.Based on the investigation results we confirmed reported issue for one of the samples.No similar customer complaints were identified.We consider this issue as an isolated incident, probably caused by material fatigue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDEX PRESSURE INFUSION C-FUSOR
Type of Device
INFUSOR, PRESSURE, FOR I.V. BAGS
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
201 west queen st.
southington CT 06489
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key13367140
MDR Text Key284532300
Report Number3012307300-2022-01912
Device Sequence Number1
Product Code KZD
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMX4810P1CZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/13/2021
Initial Date FDA Received01/27/2022
Supplement Dates Manufacturer Received05/03/2023
Supplement Dates FDA Received05/09/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-