Model Number BI70002000 |
Device Problems
Adverse Event Without Identified Device or Use Problem (2993); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/05/2022 |
Event Type
malfunction
|
Event Description
|
Medtronic received information regarding an imaging system being used during a spinal procedure.It was reported that the site got an error code when they went to take a spin.There was no reported delay and no reported impact to patient outcome.The spin was confirmed to be interrupted.After rebooting the machine the site was able to complete a successful spin.The logs were reviewed and found that the navigation system was dropping connection with the imaging system during the spin.The issue was confirmed to be on the navigation side.
|
|
Manufacturer Narrative
|
Onsite functional and visual examination was performed by a manufacturer representative.The system passed a system checkout and was determined to be operational.If information is provided in the future, a supplemental report will be issued.
|
|