Model Number 4106903
Device Problem Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/01/2021
Event Type  malfunction  
Manufacturer Narrative
One device was returned for investigation.The pinhole leak was confirmed.To try and duplicate the pinhole, the assembly and packaging process were duplicated and the failure mode could not be replicated using the tools from assembly and packaging process.Root cause could not be attributed to manufacturing process due to this.
 
Event Description
It was reported that during pre-test, a pinhole was found in the breathing bag.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX
Type of Device
ANESTHESIOLOGY
Manufacturer (Section G)
NULL
MDR Report Key13389324
MDR Text Key286859942
Report Number3012307300-2022-02143
Device Sequence Number1
Product Code CAG
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4106903
Device Catalogue NumberC45101340
Device Lot Number4106903
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/30/2021
Initial Date FDA Received01/28/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-