BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ SYRINGE; SALINE VASCULAR ACCESS FLUSH Back to Search Results
Catalog Number 306594
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/05/2022
Event Type  malfunction  
Event Description
It was reported when using the bd posiflush¿ syringe, the device experienced the stopper separating from the plunger.The following information was provided by the initial reporter.The customer stated: when the patient finished the infusion on the same day, when the nurse gave the indwelling needle to seal the tube and found that the plunger rod of the flush was separated from the stopper and could not be used, it was immediately replaced without causing adverse consequences to the patient.
 
Manufacturer Narrative
Investigation summary: a device history record review was completed by our quality engineer team for provided material number 306594 and lot number 1140615.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.Should you again experience any problems with our product we would appreciate the opportunity to conduct a thorough analysis.There are quality controls currently in place to detect this type of defect during the production process.Further action has not been determined necessary at this time.Our quality team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ SYRINGE
Type of Device
SALINE VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13392388
MDR Text Key286446872
Report Number1911916-2022-00029
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/30/2024
Device Catalogue Number306594
Device Lot Number1140615
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/11/2022
Initial Date FDA Received01/31/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-