DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/10/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device: a getinge field service engineer (fse) evaluated the cardiosave intra-aortic balloon pump (iabp) unit and was able to reproduce the reported issue.The fse determined that the three transducer readings would change when fse pressed on the connectors to the backplane board from the pneumatic interface board.To fix the issue, the fse reseated connects j19 and j20 to backplane board, replaced the screw kit, cardiosave pm schedule b and performed all functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during a routine check performed by the biomed, the cardiosave intra-aortic balloon pump (iabp) displayed an internal communication failure.There was no patient involvement, and no adverse event reported.
|
|