The initial reporter stated they received discrepant results for one patient sample tested with roche diagnostics cobas elecsys anti-tpo ver.3 on a cobas e 411 immunoassay analyzer.No incorrect results were reported outside of the laboratory.The sample was initially tested on the e411 analyzer, resulting in an anti-tpo value of 79 iu/ml (expected value = < 34 iu/ml).The result did not match the clinical picture of the patient, so the sample was sent to another laboratory for testing on a cobas 6000 e 601 module.The result from the e 601 analyzer was < 5.0 iu/ml (expected value = < 34 iu/ml) on (b)(6) 2022.The sample was repeated again on the e411 analyzer, resulting in a value of around 84 iu/ml.The anti-tpo reagent lot number was 54394103, with an expiration date of 31-mar-2022.
|