• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P DISCOVERY ELBOW; DISC HUM 4X100MM LT FLANGED C

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P DISCOVERY ELBOW; DISC HUM 4X100MM LT FLANGED C Back to Search Results
Model Number 540-00-000
Device Problems Loose or Intermittent Connection (1371); Appropriate Term/Code Not Available (3191)
Patient Problem Insufficient Information (4580)
Event Date 01/11/2022
Event Type  Injury  
Event Description
Revision surgery: implant had loosened post surgery.
 
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Manufacturer Narrative
Manufacturer narrative : the reason for this revision surgery was reported as loosening.The actual length of in-vivo for the items listed is unknown as the original surgery date was not provided or could be established.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.This investigation is limited in scope as only partial information was provided to djo surgical for review.The revised items was not returned for examination and the lot numbers were not provided.To adequately investigate this event, the lot numbers are necessary.Given the limited information, a search for an invoice (of the previous surgery) produced no results, therefore; the items removed could not be identified.If this information is submitted at a future date, this investigation will be re-evaluated.Customer complaint history of the reported items showed no present trends or on-going issues that are needing a review.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.The surgeon performed this revision to remedy the patient's condition.This complaint will be closed pending receipt of additional information.No further action is deemed necessary at this time.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISCOVERY ELBOW
Type of Device
DISC HUM 4X100MM LT FLANGED C
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
kiersten soderman
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key13418189
MDR Text Key284858992
Report Number1644408-2022-00121
Device Sequence Number1
Product Code JDC
UDI-Device Identifier00190446137610
UDI-Public(01)00190446137610
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K013042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number540-00-000
Device Catalogue Number540-34-100
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/11/2022
Initial Date FDA Received02/01/2022
Supplement Dates Manufacturer Received02/03/2022
Supplement Dates FDA Received02/21/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
540-00-000 LOT UNKNOWN.; 540-34-100 LOT UNKNOWN.
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-