ALCON RESEARCH, LLC - HOUSTON CENTURION ULTRAVIT PROBE PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752134
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Distress (2329)
Event Date 01/05/2022
Event Type  Injury  
Event Description
A customer reported the vitrectomy came off the probe, able to perform the second procedure and stress was reported.The first procedure could not be performed.No patient harm was reported.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation and the product was clearly used beyond the expiration date; therefore, the condition of the product could not be verified.A sample was not returned, and a review of the lot number provided indicated the product was used beyond its expiration date.No specific action with regard to this complaint was taken by the manufacturing site because the complaint product exceeded its expiration date and should not have been used.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION ULTRAVIT PROBE PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13440554
MDR Text Key284968115
Report Number1644019-2022-00097
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657521340
UDI-Public00380657521340
Combination Product (y/n)N
Reporter Country CodeAR
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2016
Device Catalogue Number8065752134
Device Lot Number1652923H
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/31/2022
Initial Date FDA Received02/03/2022
Supplement Dates Manufacturer Received04/27/2022
Supplement Dates FDA Received05/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/02/2014
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1