• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M Back to Search Results
Model Number BI70000027120
Device Problems No Display/Image (1183); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/13/2022
Event Type  malfunction  
Event Description
Medtronic received information regarding an imaging system being used in a ortho (tha and tka) procedure.It was reported that after the spin the scan did populate but then the navigation system was frozen.The scan populated on the mobile view station (mvs) and they had a 3d image from the spin.The site tried to push manually but it would not push.The site did not go back to ap lateral.In the original spin it looked as if the spin did not complete but it gave them an image still.The system did not finish its spin and would not manually push and unable to send.After restarting the navigation system and re-spinning everything worked fine.The site believes this is a software malfunction.Troubleshooting was performed and they checked to see if the ethernet was plugged in and out.There was less than an hour delay in the case.No impact on patient outcome.
 
Manufacturer Narrative
The manufacturer representative went to the site to test the imaging system.No issues were found while transferring images.However, the navigation system hard drive was close to full.The cleaned the hard drive and advised the site to remove data from time to time.If information is provided in the future, a supplemental report will be issued.
 
Event Description
No reported impact to the patient.
 
Manufacturer Narrative
H2) additional information was added to the event description.E) surgeons name was received.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received.It was clarified the x-ray technician was the only user in the operating room (or) when the image acquisition was performed.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13442018
MDR Text Key288808887
Report Number3004785967-2022-00072
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K092564
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI70000027120
Device Catalogue NumberBI70000027120
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/13/2022
Initial Date FDA Received02/03/2022
Supplement Dates Manufacturer Received02/03/2022
02/25/2022
Supplement Dates FDA Received02/25/2022
03/18/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age60 YR
Patient SexMale
Patient Weight82 KG
-
-