SMITH & NEPHEW, INC. REAMER HANDLE; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
Model Number 71362279 |
Device Problems
Break (1069); Mechanical Jam (2983)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/17/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference case-(b)(4).
|
|
Event Description
|
It was reported that, the reamer handle no longer locks into the power drill.As this was noticed upon field inspection, there was not patient involvement.
|
|
Manufacturer Narrative
|
B3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was fractured along the chuck end, rendering the device inoperative.This could prevent the device from being capable of connecting to the power drill.The device shows signs of extensive use.The fractured component was not returned.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.A contribution of the device to the reported event could be corroborated as the device shows signs of damage/wear.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
|
Event Description
|
It was reported that, the reamer handle no longer locks into the power drill.And during investigation it was found that the device was fractured along the chuck end, rendering the device inoperative.As this was noticed upon field inspection, there was not patient involvement.
|
|
Search Alerts/Recalls
|
|
|