• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REAMER HANDLE; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REAMER HANDLE; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71362279
Device Problems Break (1069); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference case-(b)(4).
 
Event Description
It was reported that, the reamer handle no longer locks into the power drill.As this was noticed upon field inspection, there was not patient involvement.
 
Manufacturer Narrative
B3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was fractured along the chuck end, rendering the device inoperative.This could prevent the device from being capable of connecting to the power drill.The device shows signs of extensive use.The fractured component was not returned.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.A contribution of the device to the reported event could be corroborated as the device shows signs of damage/wear.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Event Description
It was reported that, the reamer handle no longer locks into the power drill.And during investigation it was found that the device was fractured along the chuck end, rendering the device inoperative.As this was noticed upon field inspection, there was not patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAMER HANDLE
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13446382
MDR Text Key285169798
Report Number1020279-2022-00473
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010456519
UDI-Public03596010456519
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71362279
Device Catalogue Number71362279
Device Lot Number06KM09907
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2022
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/17/2022
Initial Date FDA Received02/03/2022
Supplement Dates Manufacturer Received03/02/2022
Supplement Dates FDA Received03/02/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-