It was reported to boston scientific corporation on (b)(6) 2022 that wallflex esophageal fully covered stent was to be implanted to treat a malignant stricture in the esophagus during an esophagogastroduodenoscopy (egd) with stent placement procedure performed on (b)(6) 2022.Reportedly, the patient's anatomy was tortuous and was not dilated prior to stent placement.During the procedure, the stent was difficult to deploy inside the patient.The physician removed the stent from the patient fully covered by the outer sheath and attempted to deploy the stent partially outside of the patient; however, the stent prematurely fully deployed.The procedure was completed with another wallflex esophageal stent.There were no patient complication reported as a result of this event.
|
(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|