RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 VENT, JAPAN OPT: CFLEX,AVAPS,AT+ |
Device Problem
Calibration Problem (2890)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/13/2022 |
Event Type
malfunction
|
Event Description
|
A proximal pressure sensor calibration data error alarm was reported to have occurred while the device was in use on a patient.A manufacturer sales representative confirmed the log of "proximal pressure sensor calibration data error" alarm the day following the date of occurrence.The ventilator was on a patient wearing a performatrak full face mask and due to the alarm, the hospital replaced the device with another v60 ventilator.The international service engineer evaluated the ventilator.The reported phenomenon was not duplicated despite run testing.The event log revealed that "check vent: proximal pressure sensor range error" (diagnosis code 110d) had occurred.The device was inspected but no anomaly was observed.The software was confirmed to be version 2.30.Overall checking, cleaning, run testing, and functional testing were performed on the device.
|
|
Manufacturer Narrative
|
Reporter phone number - (b)(6).
|
|
Manufacturer Narrative
|
(b)(6).
|
|
Event Description
|
A proximal pressure sensor calibration data error alarm was reported to have occurred while the device was in use on a patient.A manufacturer sales representative confirmed the log of "proximal pressure sensor calibration data error" alarm the day following the date of occurrence.The ventilator was on a patient wearing a performatrak full face mask and due to the alarm, the hospital replaced the device with another v60 ventilator.The international service engineer evaluated the ventilator.The reported phenomenon was not duplicated despite run testing.The event log revealed that "check vent: proximal pressure sensor range error" (diagnosis code 110d) had occurred.The device was inspected but no anomaly was observed.The software was confirmed to be version 2.30.Overall checking, cleaning, run testing, and functional testing were performed on the device.
|
|
Search Alerts/Recalls
|
|
|