• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. SHELL POROUS WITH CLUSTER HOLES 56 MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. SHELL POROUS WITH CLUSTER HOLES 56 MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Corroded (1131); Material Discolored (1170)
Patient Problems Pain (1994); Metal Related Pathology (4530); Muscle/Tendon Damage (4532)
Event Date 07/27/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant products: 00801803201 ¿ cocr head ¿ 60939983; 00771100900 ¿ m/l taper stem ¿ 60939440; 00625006530 ¿ bone screw- 60875832; 00625006525 ¿ bone screw - 60740475; 00630505632 ¿ liner ¿ 60918074.Medical records were reviewed by an hcp and notes revision due to pain.Pseudocapsule in the joint and corrosion at the trunnion and screw holes of the shell.Dhr was reviewed and no discrepancies were found.The root cause is unable to be determined.Multiple mdr reports were filed for this event, please see associated reports: 0002648920 - 2021 - 00310, 0001822565 - 2021 - 02832, 0002648920 - 2022 - 00022, 0002648920 - 2022 - 00023.
 
Event Description
It was reported patient underwent a right hip revision approximately 12 years post implantation due to pain.During the surgery a pseudocapsule was encountered, gross corrosion was noted at the trunnion and acetabulum components, and the gluteus medius damaged from previous surgery.The head and liner were removed and replaced without complications.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHELL POROUS WITH CLUSTER HOLES 56 MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13451338
MDR Text Key285048861
Report Number0001822565-2022-00348
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K021891
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2016
Device Model NumberN/A
Device Catalogue Number00620205622
Device Lot Number60548425
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/07/2022
Initial Date FDA Received02/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/02/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexFemale
-
-