WRIGHT MEDICAL TECHNOLOGY INC INFINITY¿ TOTAL ANKLE SYSTEM; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 33654406 |
Device Problem
Device Contaminated During Manufacture or Shipping (2969)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/06/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Analysis results are not available at the time of this report.A follow-up report will be sent when the analysis is complete.
|
|
Event Description
|
It was reported that there was a presence of hair in implant.
|
|
Manufacturer Narrative
|
The complaint is confirmed, as we have received back a open package with a hair for evaluation.Inspection of the received information/evidences are done or will be done, in the proceedings of the nc.The device inspection revealed the following: visual evaluation of the received product has shown that we have received a open carton package, furthermore also the outer sterile package is already opened.The reported hair we have found lying on the inner sterile package.The inner sterile package is still in a sealed and intact condition including the implant as it should.Also there is no hair visible inside the package.We are not able to confirm if the hair was packed and sealed in the outer package together with the inner package.If more information is provided, the case will be reassessed.
|
|
Event Description
|
It was reported that there was a presence of hair in implant.
|
|