Catalog Number 07762798016
Device Problem No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/12/2022
Event Type  malfunction  
Manufacturer Narrative
The test strips were requested for investigation and have not been received at this time.If the product is returned in the future, a follow-up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Per product labeling: "coaguchek method uses human recombinant thromboplastin.Therefore, the comparability to tests using other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastin types.However, those higher differences between thromboplastins of different (rabbit, bovine) origin are not an issue specific for coaguchek assays.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared with several other (rabbit, bovine) laboratory methods." occupation is patient/consumer.
 
Event Description
We received an allegation of a questionable % quick result for one patient tested with the coaguchek inrange meter with serial number (b)(4) compared to an unknown laboratory method.The result from the meter was 64% quick.The result from the laboratory was 50% quick.The time interval between the two tests was within 1 hour.The therapeutic range is 2.2 to 2.8 inr and the middle range is 2.5 inr.Testing is performed once a week.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key13457299
MDR Text Key287803845
Report Number1823260-2022-00307
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K170960
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 02/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/31/2023
Device Catalogue Number07762798016
Device Lot Number56561113
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/12/2022
Initial Date FDA Received02/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
BLOOD PRESSURE MEDS (2 DIFFERENT) - EVERY 2 DAYS; FALITHROM; FOLIC ACID; LUNDENEROSOL - 2X/DAY
Patient Age68 YR
Patient SexMale
-
-