• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® LIFEBAND; CARDIAC RESUSCITATOR BAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® LIFEBAND; CARDIAC RESUSCITATOR BAND Back to Search Results
Model Number MODEL 100
Device Problems Material Integrity Problem (2978); Material Twisted/Bent (2981)
Patient Problem Insufficient Information (4580)
Event Date 01/13/2022
Event Type  malfunction  
Event Description
During patient use, customer reported that the autopulse platform (serial (b)(4)) displayed fault code "16" (timeout moving to take-up position) error message.The customer immediately performed manual cpr for the remainder of the event.After the event, the customer noticed the lifeband was twisted in the sleeve and it could not be straighten, and customer decided to cut off the sleeve of the lifeband.Patient's status information was requested but the customer did not provide a response.Please see the following related mfr report: mfr # 3010617000-2022-00111 for the autopulse platform (serial (b)(4)).
 
Manufacturer Narrative
Zoll has not received the product for investigation.A follow-up report will be submitted when the product is returned and investigation has been completed.
 
Manufacturer Narrative
The reported complaint that "the autopulse lifeband (lot # 170557) had been twisted was confirmed during the visual inspection of the returned lifeband.The probable root cause of the reported complaint was likely incorrectly installed lifeband or the patient displacement/ shift on the platform during the compressions.During the visual inspection, observed the belt on band 1 was twisted, thus confirming the customer's complaint.In general, a twisted belt should not affect the functionality of the lifeband.In addition, both of the protective band sleeves were cut/torn open by the customer, unrelated to the reported complaint.It is possible, the customer wanted to verify the twisted lifeband.Functional testing could not be performed due to the observed damages on the returned lifeband.Historical complaints were reviewed for information related to the reported complaint, and there was no previous history of complaints reported for lifeband with lot# 170557.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® LIFEBAND
Type of Device
CARDIAC RESUSCITATOR BAND
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key13459386
MDR Text Key285245372
Report Number3010617000-2022-00171
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111001571
UDI-Public00849111001571
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0701-01
Device Lot Number170557
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/13/2022
Initial Date FDA Received02/04/2022
Supplement Dates Manufacturer Received03/02/2022
Supplement Dates FDA Received03/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/12/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-