Catalog Number VS-404 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Obstruction/Occlusion (2422); Thrombosis/Thrombus (4440)
|
Event Date 01/30/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Physician used the venaseal closure system to treat a segment of patient's left short saphenous vein (ssv).Transducer and hand compression were used per venaseal protocol.There were no reportable events for during or immediately post-op.Procedure was uneventful.The vein is reported to have closed.The patient presented for normal s/p venous scan of area 2 days later.The treated segment of the saphenous vein was completely non-compressible and demonstrated an echogenic lumen.There was no flow demonstrated by color-flow imaging.There was no evidence or early recanalization at the time.At this time it was noted that there was an occlusion with probable venaseal glue in the spj extending into the popliteal vein.The left small saphenous vein was occluded with no evidence of early recanalization.Endovenous ablation-induced junctional thrombus was present with occlusion of popliteal vein.Egit class:j-4.The patient was placed on eliquis.Patient was scheduled to be re-scanned 8 days post procedure.Patient status at time of report unknown.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Per account, tip was located 5cm caudal to the sapheno popliteal junction (spj).Compression was given of the ssv.Patient was put on eliquis.Correction: implant date correction.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Patient is with continuing on eliquis.Pt stated to be in no pain.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Image analysis image 1: shows the doppler gate placed over the vein (ssv) with no tracing ¿ not evident if the vein has reflux.Images 2 and 3: are the same image.Image 3 shows the measurement.(image 2 has no scale, and is an image labeled ¿measurement during treatment¿).Image 3 is the critical image and the one that is most problematic.It clearly shows a measurement of 3.37 cm for the distance of the device to the junction(deep) and this is obviously less than the required 5cm per the ifu.The report provided indicates that the procedure was done to ifu and the device was 5cm from the junction but according to this image, this cannot be proven out in the duplex images provided.Images 4 to 16: the remainder of the images appear to document via gray scale with compression/non compression side by side comparisons (typically in transverse) that the popliteal has a dvt.They also depict in long axis images of the ssv being closed with extension into the popv ¿ thus dvt.There are two color images which shows the lack of flow in the pop v ¿ in theory confirming the dvt.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|