• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. HYDROSITE AD PLUS 7.5 X 7.5CM; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. HYDROSITE AD PLUS 7.5 X 7.5CM; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number 66800674
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that when the pouch was opened, it was confirmed there was a tiny black foreign substance in the hydrosite ad plus 7.5 x 7.5cm.The treatment was completed, without delay, using a s+n back up device.Patient was not harmed as consequence of this problem.
 
Manufacturer Narrative
The device had been returned and evaluated, establishing a relationship between the device and the reported event.The returned dressing was cut and observed under microscope.The tiny black substance was embedded in pad, and suspected to be fibre of clothing.The root cause has been identified that clothing fiber dropped into foam during casting process, and the dressing escaped from on-line inspection.The manufacturing records show no evidence that the product did not meet specification at the time of manufacture.The complaint history file contains further instances of the reported events.The instructions for use contain comprehensive instructions on the safe operation and use of the device.The risk files mitigate the reported issue with no updates required.This investigation is now complete, with no corrective actions required.Operators were made aware of this complaint, and trained with gmp requirement and on-line inspection requirement.Smith and nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYDROSITE AD PLUS 7.5 X 7.5CM
Type of Device
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13478452
MDR Text Key285958465
Report Number8043484-2022-00051
Device Sequence Number1
Product Code KGX
UDI-Device Identifier05000223468989
UDI-Public5000223468989
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66800674
Device Lot Number202127
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/18/2022
Initial Date FDA Received02/08/2022
Supplement Dates Manufacturer Received03/16/2022
Supplement Dates FDA Received03/17/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-