For the investigation the logfile was analysed.On the reported date of event the case was started using monitoring mode.In monitoring mode no fresh gas is delivered to the ventilator, in the lower left corner of the display (above the gas mixer hardkeys) a no fresh gas supply message appears (no alarm).In the following the device alarmed repeatedly apnea co2, insp.O2 low (measured to 17%) and mv low.Some minutes later o2 started to increase from 17% to 56%.The user obviously used the o2 flush.Subsequently the patient was disconnected from the device.The unit was placed in standby.In the case no fresh gas is provided to the ventilator, ventilation and especially oxygenation of the patient might be restricted.By using the o2 safety flow valve, the user is able to set an adequate fresh gas flow including anesthetic agent, also if the device is switched off.O2-desaturation of patient leads to a critical situation.In general the device alerts with a high priority alarm insp.O2 low.In case of an o2 supply outage, additional alerts for o2 supply pressure low (in case the pipe- or cylinder pressure is too low) and no o2 supply (in case there is no pressure) are given visible and audible (with max.Loudness independent from user settings).The analysis of the electronic logfile has given no hints for malfunction of the device or one of its components during the case in question.The device was tested on site and no technical failures were found, which confirmes the results from the logfile analysis.The ifu provides a detailed description of the monitoring mode.
|