• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC SWITZERLAND GMBH PEDIMAG BLOOD PUMP; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC SWITZERLAND GMBH PEDIMAG BLOOD PUMP; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number 201-90052
Device Problems Mechanical Problem (1384); Infusion or Flow Problem (2964)
Patient Problems Hemolysis (1886); Inflammation (1932); Sepsis (2067); Obstruction/Occlusion (2422); Thrombosis/Thrombus (4440)
Event Date 12/28/2021
Event Type  Injury  
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted when the manufacturer's investigation is completed.
 
Event Description
It was reported that the patient was admitted on (b)(6) 2021 and was placed on pedimag support on (b)(6) 2021.The hospital team was able to visualize fibrin strands and thrombus in the pedimag circuit and a development of urosepsis and inflammation on (b)(6) 2021 lead to increased fibrin deposition.Based on frequency of transfusions and laboratory evidence of hemolysis, the patient required frequent blood transfusions.The patient¿s hematocrit (hct) 30.9 % down from 36 and required packed red blood cells (prbcs), lactate dehydrogenase (ldh) 1249 u/l, plasma free hemoglobin (hgb) 43.5 g/dl.The team decided to send the patient to the operating room (or) and have the pedimag pump and circuit exchanged.The patient tolerated the procedure well.The patient¿s neuro status was stable and patient was on 8l high-flow nasal cannula and tolerated a slow increase in nasojejunal tube feeds.The team was not concerned with any infection and listed the patient for heart transplant.
 
Manufacturer Narrative
No further information was provided.The manufacturer is closing the file on this event.
 
Manufacturer Narrative
Correction on event date.Manufacturer's investigation conclusion: the report of thrombus could not be confirmed through this investigation; no photographs were submitted for review and the device was not returned for evaluation.A direct correlation between the reported events and the pedimag device could not be determined through this evaluation.The pedimag device was not disposed of and exchanged for a new one.The centrimag blood pump instructions for use (ifu) lists thromboembolic phenomena and hemolysis as a possible side effects that may be associated with the use of the centrimag blood pump (ifu warning #3).This ifu also provides the following warnings and cautions: ifu warning #9: it is intended that systemic anticoagulation be utilized while this device is in use.Anticoagulation levels should be determined by the physician based on risks and benefits to the patient.Ifu warning #10: frequent patient and device monitoring is recommended.Ifu warning #14: do not operate the pump for more than 30 seconds in the absence of flow.The temperature within the pump will rise and increased cellular damage and clotting may result.Ifu warning #19: do not restart the pump if the pump has been stopped for more than 5 minutes without adequate anticoagulation, as the risk of thromboembolism is increased after blood has remained stagnant in the pump, extracorporeal circuit, connectors, and cannulae.Ifu caution #2: this device should only be used by persons thoroughly trained in extracorporeal circulation procedures.Ifu caution #9: monitor carefully for any signs of occlusion throughout the circuit.Ifu caution #15: always have a backup centrimag pump, console, motor, and accessories available for use.The device history record for the pedimag blood pump, lot number l06049-lb2, was reviewed and showed no deviations from manufacturing or quality assurance specifications.No further information was provided.The manufacturer is closing the file on this event.
 
Manufacturer Narrative
No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEDIMAG BLOOD PUMP
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CA CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CA CH-80 05
SZ   CH-8005
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key13481017
MDR Text Key286374515
Report Number2916596-2022-00440
Device Sequence Number1
Product Code KFM
UDI-Device Identifier07640135140634
UDI-Public07640135140634
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup,Followup
Report Date 07/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2022
Device Model Number201-90052
Device Catalogue Number201-90052
Device Lot NumberL06049-LB2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/15/2022
Initial Date FDA Received02/08/2022
Supplement Dates Manufacturer Received04/15/2022
04/28/2022
07/11/2022
Supplement Dates FDA Received04/27/2022
04/28/2022
07/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization; Life Threatening;
Patient Age2 MO
Patient SexFemale
Patient Weight5 KG
-
-