• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: A.I.D.D LONGFORD ALINITY I TSH REAGENT KIT; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

A.I.D.D LONGFORD ALINITY I TSH REAGENT KIT; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE Back to Search Results
Model Number 07P4830
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2021
Event Type  malfunction  
Manufacturer Narrative
Multiple sids were assigned to the same patient; no patient demographic information was provided.An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event Description
The customer questioned alinity i tsh results for one patient.The following was provided: sid (b)(6)- (b)(6) 2021 - initial result generated with tsh reagent lot 30385ud00 was 0.039 miu/l, new sample tested at a later date lid (b)(6) result 0.315 miu/l, new sample collected, tested and questioned however not provided by the customer.The results were reported out of the lab and were questioned by the physician.No impact to patient management was reported.
 
Manufacturer Narrative
An investigation was performed for alinity i tsh reagent lots 35259ud00.A review of tickets was performed and determined the reagent lot performs as expected for this product.A review of tracking and trending did not identify any trends for the complaint issue.Return testing was not completed as returns were not available.Accuracy testing for reagent lot 35259ud00 was completed using panels which mimic patient samples using an in-house retained kit stored at the recommended storage condition.All specifications were met indicating that the lots are performing acceptably.A review of the product quality history for the lot number using search of the corrective and preventive actions (capa) system did not identify issues associated with the customer¿s observation.A review of labeling concluded that the issue is sufficiently addressed.Based on the investigation no systemic issue or deficiency of the alinity i tsh reagent lots 35259ud00 and 30385ud00 was identified this event was additionally submitted via 3005094123-2022-00027 under a different suspected medical device lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY I TSH REAGENT KIT
Type of Device
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
433331157
MDR Report Key13481250
MDR Text Key290283737
Report Number3005094123-2022-00026
Device Sequence Number1
Product Code JLW
UDI-Device Identifier00380740131159
UDI-Public00380740131159
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
K983442
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/01/2023
Device Model Number07P4830
Device Catalogue Number07P48-30
Device Lot Number35259UD00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/31/2022
Initial Date FDA Received02/08/2022
Supplement Dates Manufacturer Received05/06/2022
Supplement Dates FDA Received05/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/06/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ALNTY I PROCESSING MODU, 03R65-01, (B)(6).; ALNTY I PROCESSING MODU, 03R65-01, (B)(6).; ALNTY I TSH RGT 1200T K, 07P48-30, (B)(6).; ALNTY I TSH RGT 1200T K, 07P48-30, (B)(6).; ALNTY I TSH RGT 1200T K, 07P48-30, (B)(6).; ALNTY I TSH RGT 1200T K, 07P48-30, (B)(6).
-
-