SMITH & NEPHEW, INC. R3 LNR MPACTOR HD II 32MM; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
Model Number 71366432 |
Device Problems
Crack (1135); Degraded (1153)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/24/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that, during set up for a thr surgery, it was noticed that two r3 lnr impactor hd ii 32mm has cracked in them due to wear.The procedure was finished with a smith and nephew back up device without delay.Since the incident occurred before the procedure, the patient was not involved.
|
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Manufacturer Narrative
|
H3, h6: the devices were not returned for evaluation and the reported event could not be confirmed.A review of complaint history of the previous 12 months revealed similar events for the listed devices, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.These devices are reusable instruments that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the devices to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the devices or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|