• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG; PNEUMO SURE XL HIGH FLOW INSUFFLATOR; INSUFFLATOR, LAPAROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG; PNEUMO SURE XL HIGH FLOW INSUFFLATOR; INSUFFLATOR, LAPAROSCOPIC Back to Search Results
Model Number 0620040610
Device Problems Pressure Problem (3012); Insufficient Information (3190)
Patient Problems Arrhythmia (1721); Respiratory Insufficiency (4462); Insufficient Information (4580)
Event Date 01/19/2022
Event Type  Injury  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event Description
It was reported that the procedure was cancelled.
 
Manufacturer Narrative
Alleged failure: read incorrect, doc said was reading incorrect/they was how ever using a 3rd party tubing and not stryker/ salesforce case number (b)(4).The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root cause/s could be a third party tubing used during the procedure and/or a loose adapter used during the procedure, as there were no findings during the functional testing of the unit.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
 
Event Description
It was reported that the procedure was cancelled.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG; PNEUMO SURE XL HIGH FLOW INSUFFLATOR
Type of Device
INSUFFLATOR, LAPAROSCOPIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key13514735
MDR Text Key285503938
Report Number0002936485-2022-00073
Device Sequence Number1
Product Code HIF
UDI-Device Identifier07613327063295
UDI-Public07613327063295
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063367
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0620040610
Device Catalogue Number0620040610
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/04/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/19/2022
Initial Date FDA Received02/11/2022
Supplement Dates Manufacturer Received01/19/2022
Supplement Dates FDA Received04/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-