• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SPEC EF PRIM HO 12/14 SZ 3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SPEC EF PRIM HO 12/14 SZ 3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71312113
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Unspecified Infection (1930); Hip Fracture (2349)
Event Date 01/24/2022
Event Type  Injury  
Event Description
It was reported that after left thr performed on (b)(6) 2021, the patient presented an infection with previous pelvic fracture.This was addressed via revisions surgery on (b)(6) 2022.Patient current status is unknown.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Manufacturer Narrative
The device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, smith and nephew has not received the device/adequate clinical documentation to fully evaluate the complaint, therefore, a thorough medical investigation cannot be rendered, nor can the clinical root cause of the reported failure be confirmed or concluded.The provided, customer feedback form supported the compliant, however, it did not contribute to determining a root cause.Should any additional relevant clinical information is later received, then the case would be re-assessed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for total hip systems revealed that primary or revision surgery has been identified as a potential complications.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the sterilization records revealed the batch was sterilized within normal parameters.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but are not limited to contamination, patient reaction, and post-operative healing issue.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPEC EF PRIM HO 12/14 SZ 3
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13517560
MDR Text Key285534171
Report Number1020279-2022-00662
Device Sequence Number1
Product Code JDI
UDI-Device Identifier03596010195586
UDI-Public03596010195586
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K970351
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/01/2015
Device Model Number71312113
Device Catalogue Number71312113
Device Lot Number05KT92875
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/24/2022
Initial Date FDA Received02/11/2022
Supplement Dates Manufacturer Received03/16/2022
Supplement Dates FDA Received03/18/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/03/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
-
-