Model Number V60
Device Problem Leak/Splash (1354)
Patient Problem Insufficient Information (4580)
Event Date 01/13/2022
Event Type  malfunction  
Event Description
A field service engineer reported a low leak ¿ co2 rebreathing risk error.It is unknown if there's patient involvement.Further information is currently pending.
 
Manufacturer Narrative
The field service engineer (fse) observed the device o2 delivery not reaching set levels.It was determined the gas delivery system was faulty.The fse replaced the gas delivery system and performed periodic maintenance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key13518909
MDR Text Key286494147
Report Number2031642-2022-00438
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/13/2022
Initial Date FDA Received02/11/2022
Supplement Dates Manufacturer Received03/24/2022
Supplement Dates FDA Received04/18/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-