• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC T.W. POWER SUPPLY; UNIT, CAUTERY, THERMAL, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC T.W. POWER SUPPLY; UNIT, CAUTERY, THERMAL, AC-POWERED Back to Search Results
Model Number T.W. POWER SUPPLY
Device Problem Failure to Deliver Energy (1211)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/20/2022
Event Type  malfunction  
Event Description
The hospital reported that during an endoscopic vein harvesting procedure using t.W.Power supply s/n (b)(4), they reported that the generator continued to beep when the hemopro2 jaws were not being activated.Customer changed out the generator to complete the case.They did not report any harm to the patient.
 
Manufacturer Narrative
Trackwise id (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
 
Manufacturer Narrative
Trackwise id # (b)(4).Analysis of production: (3331/ 213 &67)the reported device is an oem device.The certificate of conformance was reviewed for the reported serial/lot number.The vendor certifies that this device serial/lot conforms to all applicable product specifications and there were no non-conformances identified for the manufacturing batch.Historical data analysis: (4109/213 & 67) the review of the historical data indicates that no other similar complaints was reported for the same lot/serial number and reported failure mode.Communication/interviews: (4111/213/ 67) communication/interviews were performed to obtain all possible information.Trend analysis: (4110/213/67) the overall 24 month product complaint trend data for the period feb 2020 through jan 2022 was reviewed.There were no triggers identified for the review period.Device not returned: (4114/213/ 67) despite request and/ or customer indicated that the device would be returned; however, no device was returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T.W. POWER SUPPLY
Type of Device
UNIT, CAUTERY, THERMAL, AC-POWERED
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key13524600
MDR Text Key286361169
Report Number2242352-2022-00142
Device Sequence Number1
Product Code HQO
UDI-Device Identifier00607567700826
UDI-Public00607567700826
Combination Product (y/n)N
PMA/PMN Number
K052274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberT.W. POWER SUPPLY
Device Catalogue NumberVH-3010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/20/2022
Initial Date FDA Received02/14/2022
Supplement Dates Manufacturer Received04/08/2022
Supplement Dates FDA Received04/08/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/30/2018
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-