• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. CERCLAGE TENSIONER RATCHETING HANDLE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. CERCLAGE TENSIONER RATCHETING HANDLE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number CERCLAGE TENSIONER RATCHETING HANDLE
Device Problems Improper or Incorrect Procedure or Method (2017); Device Markings/Labelling Problem (2911)
Patient Problem Unspecified Tissue Injury (4559)
Event Date 11/25/2021
Event Type  malfunction  
Event Description
It was reported that during an arthro assisted latarjet surgery the patient¿s coracoid was torn through by the fibertape cerclage system and half of the glenoid was torn through due to a misunderstanding regarding the used tensioner and the scale, which led to a false adjustment of the tensioner of above 350 newtons.Two tensioners were sent out for this case which was an arthroscopic assisted latarjet cerclage.The tensioner listed (ar-7800 and ar-7801) plus the blue acl tensioner ar-1529.Ar-1529 is the tensioner that was previously always used by the arthrex employee, which max¿s out at just over 100 newtons with the unit of measure listed on there in lbf and newtons.The case went really well until tensioning where the arthrex employee didn¿t realise that the surgeon was using the black tensioner (ar- 7800 & ar-7801).This tensioner doesn¿t have the unit of measure listed on it, and the scale is in lbf.The surgeon asked how much to tension to and the arthrex employee replied that the surgeon can max the tensioner out as 100 newtons are desired, it max¿s out at that point.The surgeon then tensioned to around 80 in lbf scale, which was over 350 newtons.The patient¿s coracoid was torn through by the fibertape cerclage system and half of the glenoid was torn through.In order to remove the tapes the surgeon had to make a larger posterior incision and cut down to the glenoid to release the tapes.He then used 2 x tightrope rt¿s with 2 x dog bones over the coracoid in order to complete the procedure.The surgery was finished successfully with different devices (ar-1588rt x 2).It was not necessary to do a second surgery.No further information received.
 
Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Manufacturer Narrative
Complaint confirmed.One unpackaged ar-7800 ratcheting handle and concomitant ar-7800 tensioner were received for investigation.Visual inspection revealed discoloration across the metallic components of the ratcheting handle.Functional testing identified that the ar-7800 and ar-7801 were able to be assembled together, and the assembly ratcheted as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERCLAGE TENSIONER RATCHETING HANDLE
Type of Device
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key13526738
MDR Text Key286576623
Report Number1220246-2022-04449
Device Sequence Number1
Product Code LXH
UDI-Device Identifier00858325005176
UDI-Public00858325005176
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberCERCLAGE TENSIONER RATCHETING HANDLE
Device Catalogue NumberAR-7801
Device Lot Number104730
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/26/2022
Initial Date FDA Received02/14/2022
Supplement Dates Manufacturer Received01/26/2022
Supplement Dates FDA Received06/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/13/2017
Is the Device Single Use? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-