The catalog number identified has not been cleared in the us, but is similar to the powerport implantable port products that are cleared in the us.The pro code and 510k number for the powerport implantable port products is identified in common device name and date rec'd by mfr.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, photos were provided for review.The investigation of the reported event is currently underway.(expiry date: 01/2023).Device pending return.
|
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the powerport implantable port products that are cleared in the us.The pro code and 510k number for the powerport implantable port products is identified in d2 and g4.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one powerport isp mri port, one catheter, one cath-lock, one syringe, one j-tip guidewire loaded in a guidewire hoop, one tunneler, one 6.5fr introducer peel-apart sheath and vessel dilator were returned for evaluation.Gross visual evaluations was performed.Although sample was returned for evaluation, two electronic photos were provided for review.However, the investigation is inconclusive for the reported component missing issue as the exact circumstances at the time of the reported event are unknown.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 01/2023).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|