Manufacturer Narrative
Product analysis: the reported issues of the display was very blurry and the results were improper were verified during service.The issues were resolved by replacing the pcb assy display and the socket head screw 4-40 x 3/8, performing alignment of the actuator assembly and adjusting the contrast.Preventive maintenance was performed per specifications.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that prior to use of this act plus instrument the customer reported that the display was very blurry and the results were improper.The use of the instrument was unspecified.There was no patient involvement so no adverse patient effects occurred.
 
Manufacturer Narrative
A back-up instrument was not used for the procedure.After investigation at medtronic the complaint is confirmed for the act plus instruments reported issue with the display being blurry and the results being improper.The issue was verified during service and was resolved by replacing the pcb assy display and the socket head screw, performing alignment of the actuator assembly and adjusting the contrast.Preventive maintenance/testing was performed per specifications.No patient/clinical safety issues reported.Trends for issues with this product are reviewed at quarterly quality meetings.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACT PLUS INSTRUMENT
Type of Device
TIMER, CLOT, AUTOMATED
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key13533021
MDR Text Key286768873
Report Number2184009-2022-00010
Device Sequence Number1
Product Code GKN
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K940426
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician Assistant
Type of Report Initial,Followup
Report Date 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberACT200
Device Catalogue NumberACT200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/27/2022
Initial Date FDA Received02/15/2022
Supplement Dates Manufacturer Received02/17/2022
Supplement Dates FDA Received03/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/14/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-