BAXTER HEALTHCARE CORPORATION TUBE SET, STD VOL, STER; SYSTEM/DEVICE, PHARMACY COMPOUNDING Back to Search Results
Catalog Number H93811
Device Problems Fluid/Blood Leak (1250); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/24/2022
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the tubing of three (3) sterile repeater pump tube sets leaked.The tube sets seemed to be a bit longer in length in the section that sat in the repeater pump housing, causing a leak.This was identified during setup.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the tubing of three (3) sterile repeater pump tube sets leaked.The tube sets seemed to be a bit longer in length in the section that sat in the repeater pump housing, causing a leak.This was identified during setup.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h3, h4, h6 and h10.H4: the lot was manufactured from june 24, 2021 ¿ june 25, 2021.H10: the actual devices were not available; however, two (2) companion samples were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed in both samples and no leak was observed.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUBE SET, STD VOL, STER
Type of Device
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13538525
MDR Text Key285704086
Report Number1416980-2022-00553
Device Sequence Number1
Product Code NEP
UDI-Device Identifier00085412475721
UDI-Public(01)00085412475721
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date05/31/2024
Device Catalogue NumberH93811
Device Lot Number60305539
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/24/2022
Initial Date FDA Received02/15/2022
Supplement Dates Manufacturer Received03/10/2022
Supplement Dates FDA Received03/17/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-